# mettler-toledo
Test bench for communication with Mettler Toledo scale.
specific model ICS 425
## Modes
Mettler Toledo scales have several ways to provide weight data. For our purpose the __Server__ or __Client__ options are the best match.
Check manual section 3.6 Communication menu block (page 40)
### Server mode
The scale act as a server, on which is listening at specific IP address and port waiting for a comand.
To request weight data the following command must be send to the scale controller
````
S\r\n
````
*\r\n means CR LF*
#### Server mode configuration
2020-10-23 06:19:26 +00:00
buttons
(c) (⤺⤻) (→0←) (→T←) (i) (⇰)
2020-10-23 06:19:26 +00:00
1. keep pressed (⇰)
2. Will ask "Enter code" press (⇰)
3. Will show menu options, select Communication → Com2
````
Mode → Continuos
Param. Ethernet → According network settings
TCP mode → TCP mode → Server
→ Port → <Server port, for good luck use 3333>
→ Disc timeout → 0
````
### Client mode
The scale act as a client pushing continuos data to the server, configured as specific IP address and port
Scale is contious sending data:
````
S S 0.000 kg
S S 0.000 kg
````
#### Client mode configuration
2020-10-23 06:19:26 +00:00
buttons
(c) (⤺⤻) (→0←) (→T←) (i) (⇰)
2020-10-23 06:19:26 +00:00
1. keep pressed (⇰)
2. Will ask "Enter code" press (⇰)
3. Will show menu options, select Communication → Com2
````
Mode → Continuos
Param. Ethernet → According network settings
TCP mode → TCP mode → Client
→ Remote IP → <Client IP address>
→ Remote port → <Client remote port>
→ Connect timeout → 0
→ Disc timeout → 0
````